Fiber fusion splicer electrode rods are available for: A-80S, A-81S, FS-60E, FS-60F, FS-60A, FS-60C.
Product features: The electrode is an indispensable lossy material on the fiber fusion splicer and needs to be replaced regularly. This is also one of the routine maintenance work of the fiber fusion splicer. Therefore, judging when it is necessary to replace the electrode and mastering the replacement of the electrode is one of the skills of the user of the fusion splicer.
Precautions
When replacing the electrode, firstly, it is necessary to confirm that the fusion splicer is in the off state, remove the protective cover of the electrode chamber, loosen the screw that fixes the upper electrode, take out the upper electrode, then loosen the top wire of the lower electrode, take out the lower electrode, and secondly, Install a new electrode. The operation steps are the reverse of the method of disassembling the electrode. It is required that the gap between the two electrodes is 2.6 plus or minus 0.2 mm. After that, during the assembly process, the tip of the electrode should not be touched to prevent damage to the electrode, and the electrode is prevented from falling into the machine and causing damage. After the electrode is replaced, the arc position should be calibrated.
maintenance
As a special precision instrument, the welding machine should pay attention to avoid excessive ground vibration as much as possible. Also pay attention to waterproof and moisture proof. You can put desiccant in the cabinet and put it in a dry and ventilated place when not in use.